banner
Home

Blog

Blog

  • Bellcore GR78-CORE Test Specification Bellcore GR78-CORE Test Specification
    Aug 14, 2024
      Bellcore GR78-CORE is one of the specifications used in early surface insulation resistance measurement (such as IPC-650). The relevant precautions in this test are organized for reference of personnel who need to carry out this test, and we can also have a preliminary understanding of this specification. Test purpose: Surface Insulation Resistance Testing 1. Constant temperature and humidity test chamber: the minimum test conditions are 35°C±2°C/85%R.H., 85 ±2°C/85% R.H. 2. Ion migration measurement system: Allowing insulation resistance of the test circuit to be measured under these conditions, a power supply will be able to provide 10 Vdc / 100μA.   Test procedure: a. The test object is tested after 24 hours at 23°C (73.4° F)/50%R.H. environment b. Place limited Test patterns on an appropriate rack and keep the test circuits at least 0.5 inches apart, without obstructing air flow, and the rack in the furnace until the end of the experiment. c. Place the shelf in the center of the constant temperature and humidity test chamber, align and parallel the test board with the air flow in the chamber, and lead the line to outside of the chamber, so that the wiring is far away from the test circuit. d. Close the furnace door and set the condition to 35 ±2°C, at least 85%R.H. and allow the furnace to spend several hours stabilizing e. After 4 days, the insulation resistance will be measured and the measured value will be recorded periodically between 1 and 2,2 and 3,3 and 4, 4 and 5 using an applied voltage of 45 ~ 100 Vdc. Under the test conditions, the test is sent out the measured voltage to the circuit after 1 minute. 2 and 4 are periodically at an identical potential. And 5 periodically at opposite potentials. f. This condition only applies to transparent or translucent materials, such as solder masks and conformal coatings. g. As for multilayer printed circuit boards required for insulation resistance testing, the only normal procedure will be used for insulation resistance testing circuit products. Extra cleaning procedures are not allowed.   Related test chamber: temperature and humidity chamber Method of conformity determination: 1. After the electron migration test is completed, the test sample is removed from the test furnace, illuminated from the back and tested at 10 x magnification, and will not be found to reduce the electron migration (filamental growth) phenomenon by more than 20% between the conductors. 2. adhesives will not be used as a basis for republication when determining compliance with the 2.6.11 test method of IPC-TM-650[8] to examine appearance and surface item by item. Insulation resistance does not meet the requirements of the reasons: 1. Contaminants weld the cells like wires on insulating surface of substrate, or are dropped by water of test furnace (chamber) 2. Incompletely etched circuits will reduce insulation distance between conductors by more than permitted design requirements 3. Chafes, breaks, or significantly damages the insulation between conductors  
    Read More
  • Burn-in—Lab Companion Burn-in—Lab Companion
    Jun 12, 2024
      Burn-in is an electrical stress test that employs voltage and temperature to accelerate the electrical failure of a device. Burn-in essentially simulates the operating life of the device, since the electrical excitation applied during burn-in may mirror the worst-case bias that the device will be subjected to in the course of its use able life. Depending on the burn-in duration used, the reliability information obtained may pertain to the device's early life or its wear-out. Burn-in may be used as a reliability monitor or as a production screen to weed out potential infant mortalities from the lot.   Burn-in is usually done at 125 deg C, with electrical excitation applied to the samples. The burn-in process is facilitated by using burn-in boards (see Fig. 1) where the samples are loaded. These burn-in boards are then inserted into the burn-in oven (see Fig. 2), which supplies the necessary voltages to the samples while maintaining the oven temperature at 125 deg C. The electrical bias applied may either be static or dynamic, depending on the failure mechanism being accelerated.   Figure 1. Photo of Bare and Socket-populated Burn-in Boards The operating life cycle distribution of a population of devices may be modeled as a bath tub curve, if the failures are plotted on the y-axis against the operating life in the x-axis. The bath tub curve shows that the highest failure rates experienced by a population of devices occur during the early stage of the life cycle, or early life, and during the wear-out period of the life cycle. Between the early life and wear-out stages is a long period wherein the devices fail very sparingly.   Figure 2. Two examples of burn-in ovens Early life failure (ELF) monitor burn-in, as the name implies, is performed to screen out potential early life failures. It is conducted for a duration of 168 hours or less, and normally for only 48 hours. Electrical failures after ELF monitor burn-in are known as early life failures or infant mortality, which means that these units will fail prematurely if they were used in their normal operation. High Temperature Operating Life (HTOL) Test is the opposite of ELF monitor burn-in, testing the reliability of the samples in their wear-out phase. HTOL is conducted for a duration of 1000 hours, with intermediate read points at 168 H and 500 H. Although the electrical excitation applied to the samples are often defined in terms of voltages, failure mechanisms accelerated by current (such as electromigration) and electric fields (such as dielectric rupture) are understandably accelerated by burn-in as well.
    Read More
1 2 3 4 5 6 7
A total of7pages

leave a message

leave a message
If you are interested in our products and want to know more details,please leave a message here,we will reply you as soon as we can.
submit

home

products

WhatsApp

contact us