EC-105HTP,MTP,MTHP, High and low temperature constant temperature bath (1000L)
Project
Type
Series
HT
MT
MTH
function
Temperature occurs in a way
Dry wet bulb method
Temperature range
-20 ~ + 100 ℃
-40 ~ + 100 ℃
-40 ~ + 150 ℃
Range of temperature
Below the + 100℃
± 0.3 ℃
+Above the 101℃
―
± 0.5 ℃
Temperature distribution
Below the + 100℃
± 1.0 ℃
Above the + 101℃
―
± 2.0 ℃
The temperature drops the time
+20 ~ -20 ℃
Within 60 minutes
+20 ~ -40 ℃
Within 9 0 minutes
+20 ~ -40 ℃
Within 9 0 minutes
Temperature rise time
-20 ~ + 100 ℃
Within 45 minutes
-40 ~ + 100 ℃
Within 50 minutes
-40 ~ + 150 ℃
Within 75 minutes
The internal volume of the uterus was tested
1000L
Test room inch method (width, depth and height)
1000mm × 1000mm × 1000mm
Product inch method (width, depth and height)
1400mm × 1370mm × 1795mm
Make the material
External outfit
Test room control panel
machine room
Cold steel plate, cold steel plate beige
(Color table 2.5Y8 / 2)
Inside
Stainless steel plate (SUS304,2B polished)
Broken heat material
Test room
Hard synthetic resin
―
glass wool
door
Hard synthetic resin foam cotton, glass cotton
Project
Type
Series
HT
MT
MTH
Cooling dehumidifying device
Cooling-down method
Mechanical section shrinkage mode
Cooling medium
R404A
compressor
Output (number of units)
0.75kW (1)
1.5kW (1)
Cooling and dehumidifier
Multi-channel mixed heat sink type
The condenser
Multi-channel mixed radiator plate type (air cooling type)
Calorifier
Form
Nickel-chromium heat-resistant alloy heater
Volume
3.5kW
Blower
Form
Multi-channel mixed radiator plate type (air cooling type)
Motor capacity
40W
Controller
The temperature is set
-22.0 ~ + 102.0 ℃
-42.0 ~ + 102.0 ℃
-42.0 ~ + 152.0 ℃
Humidity is set
0 ~ 98%RH (But the temperature of the wet and dry bulb is 10-85 ℃ )
Time setting Fanny
0 ~ 999 Time 59 minutes (formula) 0 ~ 20000 Time 59 minutes (formula formula)
Set decomposition energy
Temperature 0.1℃, humidity 1% RH for 1 min
Indicate accuracy
Temperature ± 0.8℃ (tp.), humidity ± 1% RH (tp.), time ± 100 PPM
Vacation type
Value or program
Stage number
20-stage / 1 program
The number of procedures
The maximum number of incoming force (RAM) programs is 32 programs
The maximum number of internal ROM programs is 13 programs
Round-trip number
98 times maximum or unlimited
Number of round-trip repeats
Maximum 3 times
Displace the end
Pt 100Ω(at 0 ℃), gradeB(JIS C 1604-1997)
Control action
When splitting the PID action
Endovirus function
Early delivery function, standby function, setting value maintenance function, power outage protection function,
Power action selection function, maintenance function, transportation round-trip function,
Time delivery function, time signal output function, overrising and overcooling prevention function,
Abnormal representation function, external alarm output function, setting paradigm representation function,
Transport type selection function, the calculation time represents the function, the slot lamp lamp function
Project
Type
Series
HT
MT
MTH
Control panel
Equipment machine
LCD operating panel (type contact panel),
Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal,
Time signal output terminal, power cord connector
Protective device
Refrigerating cycle
Overload protection device, high blocking device
Calorifier
Temperature over-rise protection device, temperature fuse
Blower
Overload protection device
Control panel
Leakage breaker for power supply, fuse (for heater, humidifier),
Fuse (for operating loop), temperature rise protection device (for testing),
Temperature rise overcooling prevention device (test material, in microcomputer)
Offproducts (sets)
House receiving (4), house board (2), operation instructions (1)
Equipment products
Adventitia
hard borosilicate glass 270mm× 190mm
2
Cable hole
内径 50mm
1
The trough inside the lamp
AC100V 15W White hot ball
2
Wheel
4
Horizontal adjustment
4
Electrovirus characteristics
Source *
AC three-phase 380V 50Hz
Maximum load current
13A
15A
Capacity of the leakage breaker for the power supply
25 A
Sensory current 30mA
Power distribution thickness
8mm2
14mm2
Rubber insulation hose
Coarseness of grounding wire
3.5mm2
5.5mm2
Tubing
drain-pipe *
PT1/2
Product weight
470kg
540kg
EC-35EXT, Superior constant temperature bath (306L)
Project
Type
Series
EXT
Function
Temperature occurs in a way
Dry wet bulb method
Temperature range
-70 ~ +150 ℃
Range of temperature
Below the + 100℃
±0.3 ℃
Above the + 101℃
±0.5 ℃
Temperature distribution
Below the + 100℃
±0. 7 ℃
Above the + 101℃
±1.0 ℃
The temperature drops the time
+125 ~-55 ℃
Within 18 points (10℃ / point average temperature change)
Temperature rise time
-55 ~+125 ℃
Within 18 minutes (10℃ / minute)
The internal volume of the uterus was tested
306L
Test room inch method (width, depth and height)
630mm × 540mm × 900mm
Product inch method (width, depth and height)
1100mm × 1960mm × 1900mm
Make the material
External outfit
Test room control panel
machine room
Cold interductile steel plate is dark gray
Inside
Stainless steel plate (SUS304,2B polished)
Broken heat material
Test room
Hard synthetic resin
door
Hard synthetic resin foam cotton, glass cotton
Project
Type
Series
EXT
Cooling dehumidifying device
Cooling-down method
Mechanical section shrinkage and freezing mode and binary freezing mode
Cooling medium;coolant
Single segment side
R 404A
Binary high temperature / low temperature side
R 404A / R23
Cooling and dehumidifier
Multi-channel mixed heat sink type
The condenser
(water-cooled)
Calorifier
Form
Nickel-chromium heat-resistant alloy heater
Blower
Form
Stir fan
Controller
The temperature is set
-72.0 ~ + 152.0 ℃
Time setting Fanny
0 ~ 999 Time 59 minutes (formula) 0 ~ 20000 Time 59 minutes (formula formula)
Set decomposition energy
Temperature was 0.1℃ for 1 min
Indicate accuracy
Temperature ± 0.8℃ (typ.), time ± 100 PPM
Vacation type
Value or program
Stage number
20-stage / 1 program
The number of procedures
The maximum number of incoming force (RAM) programs is 32 programs
The maximum number of internal ROM programs is 13 programs式
Round-trip number
Max. 98, or unlimited
Number of round-trip repeats
Maximum 3 times
Displace the end
Pt 100Ω ( at 0 ℃ ),grade ( JIS C 1604-1997 )
Control action
When splitting the PID action
Endovirus function
Early delivery function, standby function, setting value maintenance function, power outage protection function,
Power action selection function, maintenance function, transportation round-trip function,
Time delivery function, time signal output function, overrising and overcooling prevention function,
Abnormal representation function, external alarm output function, setting paradigm representation function,
Transport type selection function, the calculation time represents the function, the slot lamp lamp function
Project
Type
Series
EXH
Control panel
Equipment machine
LCD operating panel (type contact panel),
Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal,
Time signal output terminal, power cord connector
Protective device
Refrigerating cycle
Overload protection device, high blocking device
Calorifier
Temperature over-rise protection device, temperature fuse
Blower
Overload protection device
Control panel
Leakage breaker for power supply, fuse (heater,),
Fuse (for operating loop), temperature rise protection device (for testing),
Temperature rise overcooling prevention device (test material, in microcomputer)
Pay belongs to the product
Test material shed shed by * 8
Stainless steel Shshed (2), shed (4)
Fuse
Operating loop Protection Fuses (2)
Operating specification
( 1 )
Else
Bolus (Cable hole: 1)
Equipment products
Adventitia
Heat-resistant glass: 270mm: 190mm
1
Cable hole
Inner diameter of 50mm
1
The trough inside the lamp
AC100V 15W White hot ball
1
Wheel
6
Horizontal adjustment
6
Electrovirus characteristics
Power supply is * 5.1
AC Three-phase 380V 50Hz
Maximum load current
60A
Capacity of the leakage breaker for the power supply
80A
Sensory current 30mA
Power distribution thickness
60mm2
Rubber insulation hose
Coarseness of grounding wire
14mm2
Cooling water at * 5.3
Water yield
5000 L /h (When the cooling water inlet temperature is 32℃)
water pressure
0.1 ~ 0.5MPa
Side pipe diameter of the device
PT1 1/4
Tubing
Drain-pipe * 5.4
PT1/2
Product weight
700kg