TC series high and low temperature humidity climate test chamber
Product Features
• Energy saving, space saving, noise reduction, perfect pursuit of temperature and humidity test chamber
The concentrated application of numerous cutting-edge technologies has created the C-series high and low temperature and humidity test chamber.
• More convenient for user operation, with a large observation window, USB recording function, RS485 (RS232) communication interface, fully automatic water replenishment system, 7 "true color touch screen controller, etc
• More convenient and flexible testing, with one lead hole on each side with a diameter of ø 50mm (optional ø 100mm lead hole) as standard.
• Easy to maintain and upkeep, concealed electrical control cabinets, front mounted electrical control boxes, and other comprehensive protection and alarm prompts, as well as maintenance and upkeep introductions, making it more convenient for operators and maintenance technicians to maintain and upkeep equipment
Efficient and energy-saving, utilizing low-temperature throttling technology and high-performance refrigeration components, the equipment not only saves energy and water consumption for humidification, but also has higher reliability
The double-layer silicone rubber sealing ring and self-locking device effectively seal the door and testing room
The adoption of new refrigeration technologies such as more precise control and low-temperature throttling technology has improved the stability and uniformity of the equipment. The C series products have reached a new level in testing, providing you with more accurate and reliable product testing results.
• Save space, building area is extremely valuable in today's laboratory environment. The ultra-thin design of the C-series climate test chamber provides them with a narrower footprint without sacrificing valuable workspace. The optional left and right doors can adapt to different laboratory layouts, maximizing the use of ground space, while the bottom casters also increase the mobility and convenience of the equipment.
• Personalization, users can choose stainless steel casing at no additional cost, and the outer surface of the test chamber can be painted blue or gray.
Low noise, vortex compressor technology makes the refrigeration efficiency of the low-temperature system of the "C" series enclosure higher, with lower noise and more durable compressors;
Environmentally friendly, like all environmental test chambers, only HFC refrigerants are used.
• It has functions such as Chinese English conversion, date and time adjustment.
Through continuous research and practice, high-performance environmental testing equipment that meets international standards has been developed. We are constantly exploring, improving product quality, and seeking technological innovation.
Safety protection device
1. Leakage circuit breaker for power supply;
2. Anti dry burning device;
3. Compressor high pressure protection;
4. Compressor overheat protection;
5. Compressor overcurrent protection;
6. Solid state relay overload and short circuit protection circuit breaker;
7. Overload and short circuit protection fuse for heating control circuit;
8. Overload and short circuit protection fuse for humidification control circuit;
9. Overload relay for blower used for condenser heat dissipation;
10. Independent ultra-high temperature and ultra-low temperature protection;
11. Overheating of the mixing fan;
12. Reverse phase protection;
13. Phase loss protection;
14. Water shortage protection;
16. Humidifier dry burning protection;
17. Humidifier liquid level protection;
18. Laboratory ultra-high temperature and ultra-low temperature protection.
Optional features for all models
• Power socket
• Liquid nitrogen refrigeration
• Nitrogen blowing system
Adjustable stainless steel shelves
• Temperature protection for test samples
Dual channel recorder
• Pure water purifier softens water
Replace the ink cartridge of the water purifier
Dry air blowing system
• Water Recirculation System
• Water cooling unit
Hand hole with silicone gloves
Temperature and humidity control range diagram (tested under no-load conditions at an ambient temperature of 25 ℃)
Temperature and humidity chart
1. Standard Area - Black
2. Dew point temperature can be controlled up to -20 ° C (optional: compressed air dryer+capacitive humidity sensing device) - with dry air blowing
3. Long term use of high temperature and high humidity - choose special sensors
technical parameter
Model |
TC40 |
TC70 |
||||||
Test volume L |
200 |
336 |
504 |
1000 |
200 |
336 |
504 |
1000 |
Internal dimensions Depth mm |
550 |
800 |
700 |
1000 |
550 |
800 |
700 |
1000 |
Width mm |
600 |
600 |
800 |
1000 |
600 |
600 |
800 |
1000 |
Height mm |
700 |
700 |
900 |
1000 |
700 |
700 |
900 |
1000 |
Dimensions Depth mm |
1480 |
1730 |
1630 |
2000 |
1480 |
1730 |
1630 |
2000 |
Width mm |
850 |
850 |
1100 |
1250 |
850 |
850 |
1100 |
1250 |
Height mm |
1600 |
1600 |
2000 |
2150 |
1600 |
1600 |
2000 |
2150 |
Temperature test |
||||||||
Temperature range |
-40℃~150℃ |
-70℃~150℃ |
||||||
Temperature fluctuation |
≤1℃ |
|||||||
Temperature gradient |
≤2.0℃ |
|||||||
Temperature deviation |
±2.0℃ |
|||||||
According to IEC 60068-3-6 temperature cooling rate |
3.5 |
3 |
3 |
3 |
2.5 |
2 |
2 |
2 |
According to IEC 60068-3-6 temperature heating rate |
4.5 |
4 |
4 |
4 |
4.5 |
4 |
4 |
4 |
Climate testing |
||||||||
Temperature range |
+10℃~+85℃ |
|||||||
Humidity range |
10~98%RH |
10~98%RH |
||||||
Temperature fluctuation range of |
±0.1℃~±0.3℃ |
|||||||
Temperature deviation |
±0.5℃~±1℃ |
|||||||
Humidity deviation |
+2/-3%RH(when below 75% RH, the deviation is ± 5% RH) |
|||||||
Power supply |
380V,50HZ |
Note:
1. The above product dimensions do not include protruding parts;
2. The data collection of temperature and humidity values is measured by sensors under the condition of no load in the testing room. After the actual measured temperature value stabilizes to the set value, the measurement of temperature fluctuation is carried out. The number of temperature uniformity is based on a 9-point test and is measured after stabilization. The uniformity depends on the unit and operating environment, and may have slight differences.
3. All the above test data were measured at room temperature of 25 ℃.
4. The product overview provided in this document is for reference only and is not a related suggestion or recommendation, nor is it part of any contract. Due to the continuous updates of our company's products, we reserve the right to change technical specifications without prior notice. If the existing specifications are not suitable for your requirements, please ask us for customization.