banner
Home

Constant Temperature Test Chamber

Constant Temperature Test Chamber

  • Basic troubleshooting methods for high and low temperature test chambers Basic troubleshooting methods for high and low temperature test chambers
    Oct 16, 2024
    Basic troubleshooting methods for high and low temperature test chambers: 1、 High and low temperature testing equipment. In high temperature testing, if the temperature change does not reach the test temperature value, the electrical system can be checked and the faults can be eliminated one by one. If the temperature rises slowly, you need to check the air circulation system to see if the regulating baffle of the air circulation is open normally. Otherwise, check the motor of the air circulation Is the operation normal. If the temperature overshoot is severe, it is necessary to adjust the PID setting parameters. If the temperature rises directly and is protected against overheating, the controller will malfunction and the control instrument must be replaced. 2、 When the high and low temperature test equipment suddenly malfunctions during the test operation, the corresponding fault display prompt and audible alarm prompt will appear on the control instrument. The operator can quickly identify which type of fault it belongs to by referring to the troubleshooting chapter in the operation and use of the equipment, and then ask professional personnel to quickly troubleshoot it to ensure the normal progress of the experiment. Other environmental testing equipment may experience other phenomena during use, so it is necessary to analyze and eliminate them specifically. Regular maintenance and upkeep of environmental testing equipment, regular cleaning of the condenser in the refrigeration system, lubrication of moving parts according to the instructions, and regular maintenance and inspection of the electrical control system are essential tasks 3、 If the low temperature of the high and low temperature testing instrument cannot meet the test indicators, then you need to observe the temperature changes, whether the temperature drops very slowly or there is a trend of temperature recovery after reaching a certain value. The former needs to check whether the working chamber is dried before conducting the low temperature test, so that the working chamber can be kept dry before putting the test sample into the working chamber for further testing. If there are too many test samples placed in the working chamber, which prevent the air in the working chamber from fully circulating, after ruling out the above reasons, you need to consider whether it is a fault in the refrigeration system. In this case, you need to hire professional personnel from the Lab Companion manufacturer for maintenance. The latter phenomenon is caused by poor usage environment of the equipment. The temperature and location of the equipment placement (distance between the box and the wall) must meet the requirements (as specified in the equipment operation instructions). At present, the company's main products include: high and low temperature test chambers, rapid temperature change test chambers, constant temperature and humidity test chambers, and high and low temperature impact test chambers.
    Read More
  • Service conditions for high, low temperature, and low pressure test chambers Service conditions for high, low temperature, and low pressure test chambers
    Oct 15, 2024
    Service conditions for high, low temperature, and low pressure test chambers One of the usage conditions for high, low temperature, and low pressure test chambers: environmental conditions a、 Temperature: 15 ℃~35 ℃; b、 Relative humidity: not exceeding 85%; c、 Atmospheric pressure: 80kPa~106kPa d、 There is no strong vibration or corrosive gas in the surrounding area; e、 No direct sunlight exposure or direct radiation from other cold or heat sources; f、 There is no strong airflow around, and when the surrounding air needs to be forced to flow, the airflow should not be directly blown onto the box; g、 The influence of magnetic field on the control circuit of the interference free test box in the surrounding area; h、 There is no high concentration of dust or corrosive substances in the surrounding area. Condition 2 for the use of high, low temperature, and low pressure test chambers: Power supply conditions a、 AC voltage: 220V ± 22V or 380V ± 38V; b、 Frequency: 50HZ ± 0.5HZ Condition Three for the Use of High, Low Temperature, and Low Pressure Test Chambers: Water Supply Conditions It is advisable to use tap water or circulating water that meets the following conditions: a、 Water temperature: not higher than 30 ℃; b、 Water pressure: 0.1MPa~0.3MPa; c、 Water quality: meets industrial water standards. Condition 4 for the use of high, low temperature, and low pressure test chambers: Test load conditions The load of the test chamber should meet the following conditions every week: a、 The total mass of the load shall not exceed 80KG per cubic meter within the working chamber volume b、 The total volume of the load shall not exceed 5/1 of the working chamber volume c、 On any cross-section perpendicular to the prevailing wind direction, the sum of the load areas should not exceed 3/1 of the cross-sectional area of the working chamber at that location, and the load should not obstruct the flow of airflow when placed. Dear customer: Our company has products such as rapid temperature change test chambers, UV accelerated weather resistance testing machines, and temperature and humidity control chambers. You can call our service hotline through our website to learn more about our products. Our pursuit is endless, and we welcome new and old customers to choose their favorite products with confidence. We will be dedicated to serving you!
    Read More
  • Precision Oven Manual Precision Oven Manual
    Oct 15, 2024
    Precision Oven Manual Precision ovens are suitable for semiconductor devices in the electronic industry, curing and aging of electronic components, high-temperature precision testing of plastic and rubber, molding processes for telephone handle wires, as well as experimental or workshop production lines in higher education research institutions and industrial and mining enterprises that require high product temperatures. This instrument is equipped with a two-level temperature control system, dual protection, automatic cut-off for overheating, safe and reliable. The column alarm device has a temperature rise and constant temperature light display. When using this instrument in large quantities in the production workshop, which instrument has reached the constant temperature requirement and which one is still in the heating state can be clearly seen. The instrument liner is made of high-quality mirror stainless steel, the outer shell is sprayed with plastic, and a safety door lock is installed. The front door adopts a high-temperature resistant glass observation window, which can observe the condition of the test piece inside the box at any time. Dear customer: Our company has products such as rapid temperature change test chambers, UV accelerated weather resistance testing machines, and temperature and humidity control chambers. You can call our service hotline through our website to learn more about our products. Our pursuit is endless, and we welcome new and old customers to choose their favorite products with confidence. We will be dedicated to serving you!
    Read More
  • A Brief Analysis of the Five Characteristics of Walk in Laboratories A Brief Analysis of the Five Characteristics of Walk in Laboratories
    Oct 15, 2024
    A Brief Analysis of the Five Characteristics of Walk in Laboratories The walk-in laboratory has been upgraded on the basis of the original walk-in laboratory, with the characteristics of large testing space and operators being able to operate the test products in the laboratory, providing conditions for temperature and humidity environment testing for industrial manufacturers' batch or large parts, semi-finished products, and finished products. Adopting advanced Chinese LCD display screen touch screen, various complex program settings can be carried out. The program settings adopt dialogue mode, and the operation is simple and fast. It can achieve automatic operation of the refrigeration machine, maximizing automation, and can be equipped with LAN communication interfaces for users to remotely process and centrally control. It can record temperature and temperature parameters for 90 days, and is equipped with a paperless recorder. 5 characteristics of walk-in laboratory 1. Having an extremely wide temperature and humidity control range, it can meet various needs of users. By adopting a unique balanced temperature and humidity control method, a safe and precise temperature and humidity environment can be achieved. Having stable and balanced heating and humidification performance, it can achieve high-precision and highly stable temperature and humidity control. 2. Equipped with high-precision intelligent temperature regulators, temperature and humidity are displayed using LED digital display. Optional temperature and humidity recorder. 3. The refrigeration circuit is automatically selected, and the automatic control device has the performance of automatically selecting and operating the refrigeration circuit according to the set value of temperature, achieving direct start of the refrigeration machine and direct cooling under high temperature conditions. 4. The inner door is equipped with a large observation window, which facilitates the observation of the experimental status of the test samples. 5. Equipped with advanced safety and protection devices - residual current circuit breaker, over temperature protector, phase loss protector, and water cut-off protector. Dear customer: Our company has products such as rapid temperature change test chambers, UV accelerated weather resistance testing machines, and temperature and humidity control chambers. You can call our service hotline through our website to learn more about our products. Our pursuit is endless, and we welcome new and old customers to choose their favorite products with confidence. We will be dedicated to serving you!
    Read More
  • User selection environment test box must read User selection environment test box must read
    Oct 15, 2024
    User selection environment test box must read 1、 Equipment selection criteria There is currently no exact number of natural environmental factors and induced environmental factors that exist on the surface of the Earth and in the atmosphere, among which there are no less than a dozen factors that have a significant impact on the use and lifespan of engineering products (equipment). Engineers engaged in the study of environmental conditions for engineering products have compiled and summarized the environmental conditions that exist in nature and are induced by human activities into a series of testing standards and specifications to guide the environmental and reliability testing of engineering products. For example, GJB150- the National Military Standard of the People's Republic of China for Environmental Testing of Military Equipment, and GB2423- the National Standard of the People's Republic of China for Environmental Testing of Electrical and Electronic Products, which guides environmental testing of electrical and electronic products. Therefore, the main basis for selecting environmental and reliability testing equipment is the testing specifications and standards of engineering products. Secondly, in order to standardize the tolerance of environmental testing conditions in experimental equipment and ensure the control accuracy of environmental parameters, national technical supervision agencies and various industrial departments have also formulated a series of calibration regulations for environmental testing equipment and detection instruments. Such as the national standard GB5170 of the People's Republic of China "Basic Parameter Calibration Method for Environmental Testing Equipment of Electrical and Electronic Products", and JJG190-89 "Trial Calibration Regulations for Electric Vibration Test Stand System" issued and implemented by the State Administration of Technical Supervision. These verification regulations are also an important basis for selecting environmental and reliability testing equipment. Testing equipment that does not meet the requirements of these verification regulations is not allowed to be put into use. 2、 Basic principles for equipment selection The selection of environmental and reliability testing equipment should follow the following five basic principles: 1. Reproducibility of environmental conditions It is impossible to fully and accurately reproduce the environmental conditions that exist in nature in the laboratory. However, within a certain tolerance range, people can accurately and approximately simulate the external environmental conditions that engineering products undergo during use, storage, transportation, and other processes. This passage can be summarized in engineering language as follows: "The environmental conditions (including platform environment) created by the testing equipment around the tested product should meet the requirements of the environmental conditions and their tolerances specified in the product testing specifications. The temperature box used for military product testing should not only meet the requirements of the national military standards GJB150.3-86 and GJB150.4-86 for different uniformity and temperature control accuracy. Only in this way can the reproducibility of environmental conditions be ensured in environmental testing. 2. Repeatability of environmental conditions An environmental testing equipment may be used for multiple tests of the same type of product, and a tested engineering product may also be tested in different environmental testing equipment. In order to ensure the comparability of test results obtained for the same product under the same environmental testing conditions specified in the testing specifications, it is necessary to require the environmental conditions provided by the environmental testing equipment to be reproducible. This means that the stress levels (such as thermal stress, vibration stress, electrical stress, etc.) applied by environmental testing equipment to the tested product are consistent with the requirements of the same testing specification. The repeatability of environmental conditions provided by environmental testing equipment is guaranteed by the national metrological verification department after passing the verification according to the verification regulations formulated by the national technical supervision agency. Therefore, it is necessary to require environmental testing equipment to meet the requirements of various technical indicators and accuracy indicators in the calibration regulations, and to not exceed the time limit specified in the calibration cycle in terms of usage time. If a very common electric vibration table is used, in addition to meeting technical indicators such as excitation force, frequency range, and load capacity, it must also meet the requirements of precision indicators such as lateral vibration ratio, table acceleration uniformity, and harmonic distortion specified in the calibration regulations. Moreover, the service life after each calibration is two years, and after two years, it must be re calibrated and qualified before being put into use. 3. Measurability of environmental condition parameters The environmental conditions provided by any environmental testing equipment must be observable and controllable. This is not only to limit the environmental parameters within a certain tolerance range and ensure the reproducibility and repeatability of the test conditions, but also necessary for the safety of product testing, in order to prevent damage to the tested product caused by uncontrolled environmental conditions and unnecessary losses. At present, various experimental standards generally require that the accuracy of parameter testing should not be less than one-third of the allowable error under experimental conditions. 4. Exclusion of environmental testing conditions Every time an environmental or reliability test is conducted, there are strict regulations on the category, magnitude, and tolerance of environmental factors, and non test required environmental factors are excluded from penetrating into it, in order to provide a definite basis for judging and analyzing product failure and fault modes during or after the test. Therefore, it is required that environmental testing equipment not only provide the specified environmental conditions, but also not allow any other environmental stress interference to be added to the tested product. As defined in the verification regulations for electric vibration tables, the table leakage magnetic flux, acceleration signal-to-noise ratio, and total root mean square value ratio of in band and out of band acceleration. The accuracy indicators such as random signal verification and harmonic distortion are all established as verification items to ensure the uniqueness of environmental testing conditions. 5. Safety and reliability of experimental equipment Environmental testing, especially reliability testing, has a long testing cycle and sometimes targets high-value military products. During the testing process, testing personnel often need to operate, inspect or test around the site. Therefore, it is required that environmental testing equipment must have the characteristics of safe operation, convenient operation, reliable use, and long working life to ensure the normal progress of the testing itself. The various protection, alarm measures, and safety interlock devices of the testing equipment should be complete and reliable to ensure the safety and reliability of the testing personnel, the tested products, and the testing equipment itself. 3、 Selection of Temperature and Humidity Chamber 1. Selection of Capacity When placing the test product (components, assemblies, parts or whole machine) into a climate chamber for testing, in order to ensure that the atmosphere around the test product can meet the environmental testing conditions specified in the test specifications, the working dimensions of the climate chamber and the overall dimensions of the test product should follow the following regulations: a) The volume of the tested product (W × D × H) shall not exceed (20-35)% of the effective working space of the test chamber (20% is recommended). For products that generate heat during testing, it is recommended to use no more than 10%. b) The ratio of the windward cross-sectional area of the tested product to the total area of the test chamber on that section shall not exceed (35-50)% (35% is recommended). c) The distance between the outer surface of the tested product and the wall of the test chamber should be kept at least 100-150mm (recommended 150mm). The above three provisions are actually interdependent and unified. Taking a 1 cubic meter cube box as an example, an area ratio of 1: (0.35-0.5) is equivalent to a volume ratio of 1: (0.207-0.354). A distance of 100-150mm from the box wall is equivalent to a volume ratio of 1: (0.343-0.512). In summary, the working chamber volume of the climate environment test chamber should be at least 3-5 times the external volume of the tested product. The reasons for making such regulations are as follows: After the test piece is placed in the box, it occupies the smooth channel, and narrowing the channel will lead to an increase in airflow velocity. Accelerate the heat exchange between the airflow and the test piece. This is inconsistent with the reproduction of environmental conditions, as relevant standards stipulate that the air flow velocity around the test specimen in the test chamber should not exceed 1.7m/s for temperature environmental tests, in order to prevent the test specimen and the surrounding atmosphere from generating heat conduction that is not in line with reality. When unloaded, the average wind speed inside the test chamber is 0.6-0.8m/s, not exceeding 1m/s. When the space and area ratio specified in points a) and b) are met, the wind speed in the flow field may increase by (50-100)%, with an average maximum wind speed of (1-1.7) m/s. Meet the requirements specified in the standards. If the volume or windward cross-sectional area of the test piece is increased without restrictions during the experiment, the actual airflow speed during the test will exceed the maximum wind speed specified in the test standard, and the validity of the test results will be questioned. The accuracy indicators of environmental parameters in the working chamber of the climate chamber, such as temperature, humidity, salt spray settling rate, etc., are all measured under no-load conditions. Once the test piece is placed, it will have an impact on the uniformity of the environmental parameters in the working chamber of the test chamber. The larger the space occupied by the test piece, the more severe this impact will be. Experimental data shows that the temperature difference between the windward and leeward sides in the flow field can reach 3-8 ℃, and in severe cases, it can be as high as 10 ℃ or more. Therefore, it is necessary to meet the requirements of a] and b] as much as possible to ensure the uniformity of environmental parameters around the tested product. According to the principle of heat conduction, the temperature of the airflow near the box wall is usually 2-3 ℃ different from the temperature at the center of the flow field, and may even reach 5 ℃ at the upper and lower limits of high and low temperatures. The temperature of the box wall differs from the temperature of the flow field near the box wall by 2-3 ℃ (depending on the structure and material of the box wall). The greater the difference between the test temperature and the external atmospheric environment, the greater the temperature difference. Therefore, the space within a distance of 100-150mm from the box wall is unusable. 2. Selection of temperature range At present, the range of temperature test chambers abroad is generally -73 to+177 ℃, or -70 to+180 ℃. Most domestic manufacturers generally operate at -80 to+130 ℃, -60 to+130 ℃, -40 to+130 ℃, and there are also high temperatures up to 150 ℃. These temperature ranges can usually meet the temperature testing needs of the vast majority of military and civilian products in China. Unless there are special requirements, such as products installed near heat sources such as engines, the upper temperature limit should not be blindly increased. Because the higher the upper limit temperature, the greater the temperature difference between the inside and outside of the box, and the poorer the uniformity of the flow field inside the box. The smaller the available studio size. On the other hand, the higher the upper limit temperature value, the higher the heat resistance requirements for insulation materials (such as glass wool) in the interlayer of the box wall. The higher the requirement for the sealing of the box, the higher the production cost of the box. 3. Selection of humidity range The humidity indicators given by domestic and foreign environmental test chambers are mostly 20-98% RH or 30-98% RH. If the humid heat test chamber does not have a dehumidification system, the humidity range is 60-98%. This type of test chamber can only perform high humidity tests, but its price is much lower. It is worth noting that the corresponding temperature range or minimum dew point temperature should be indicated after the humidity index. Because relative humidity is directly related to temperature, for the same absolute humidity, the higher the temperature, the lower the relative humidity. For example, if the absolute humidity is 5g/Kg (referring to 5g of water vapor in 1kg of dry air), when the temperature is 29 ℃, the relative humidity is 20% RH, and when the temperature is 6 ℃, the relative humidity is 90% RH. When the temperature drops below 4 ℃ and the relative humidity exceeds 100%, condensation will occur inside the box. To achieve high temperature and high humidity, simply spray steam or atomized water droplets into the air of the box for humidification. Low temperature and humidity are relatively difficult to control because the absolute humidity at this time is very low, sometimes much lower than the absolute humidity in the atmosphere. It is necessary to dehumidify the air flowing inside the box to make it dry. At present, the vast majority of temperature and humidity chambers both domestically and internationally adopt the principle of refrigeration and dehumidification, which involves adding a set of refrigeration light pipes to the air conditioning room of the chamber. When humid air passes through a cold pipe, its relative humidity will reach 100% RH, as the air saturates and condenses on the light pipe, making the air drier. This dehumidification method theoretically can reach dew point temperatures below zero degrees, but when the surface temperature of the cold spot reaches 0 ℃, the water droplets condensed on the surface of the light pipe will freeze, affecting the heat exchange on the surface of the light pipe and reducing the dehumidification capacity. Also, because the box cannot be completely sealed, humid air from the atmosphere will seep into the box, causing the dew point temperature to rise. On the other hand, the moist air flowing between the light tubes only reaches saturation at the moment of contact with the light tubes (cold spots) and releases water vapor, so this dehumidification method is difficult to keep the dew point temperature inside the box below 0 ℃. The actual minimum dew point temperature achieved is 5-7 ℃. A dew point temperature of 5 ℃ is equivalent to an absolute moisture content of 0.0055g/Kg, corresponding to a relative humidity of 20% RH at a temperature of 30 ℃. If a temperature of 20 ℃ and a relative humidity of 20% RH are required, with a dew point temperature of -3 ℃, it is difficult to use refrigeration for dehumidification, and an air drying system must be selected to achieve it. 4. Selection of control mode There are two types of temperature and humidity test chambers: constant test chamber and alternating test chamber. The ordinary high and low temperature test chamber generally refers to a constant high and low temperature test chamber, which is controlled by setting a target temperature and has the ability to automatically maintain a constant temperature to the target temperature point. The control method of the constant temperature and humidity test chamber is also similar, setting a target temperature and humidity point, and the test chamber has the ability to automatically maintain a constant temperature to the target temperature and humidity point. The high and low temperature alternating test chamber has one or more programs for setting high and low temperature changes and cycles. The test chamber has the ability to complete the test process according to the preset curve, and can accurately control the heating and cooling rates within the maximum heating and cooling rate capability range, that is, the heating and cooling rates can be controlled according to the slope of the set curve. Similarly, the high and low temperature alternating humidity test chamber also has preset temperature and humidity curves, and the ability to control them according to the preset. Of course, alternating test chambers have the function of constant test chambers, but the manufacturing cost of alternating test chambers is relatively high because they need to be equipped with curve automatic recording devices, program controllers, and solve problems such as turning on the refrigeration machine when the temperature in the working room is high. Therefore, the price of alternating test chambers is generally more than 20% higher than that of constant test chambers. Therefore, we should take the need for experimental methods as the starting point and choose a constant test chamber or an alternating test chamber. 5. Selection of variable temperature rate Ordinary high and low temperature test chambers do not have a cooling rate indicator, and the time from the ambient temperature to the nominal lowest temperature is generally 90-120 minutes. The high and low temperature alternating test chamber, as well as the high and low temperature alternating wet heat test chamber, both have temperature change speed requirements. The temperature change speed is generally required to be 1 ℃/min, and the speed can be adjusted within this speed range. The rapid temperature change test chamber has a fast temperature change rate, with heating and cooling rates ranging from 3 ℃/min to 15 ℃/min. In certain temperature ranges, the heating and cooling rates can even reach over 30 ℃/min. The temperature range of various specifications and speeds of rapid temperature change test chambers is generally the same, that is, -60 to+130 ℃. However, the temperature range for assessing the cooling rate is not the same. According to different test requirements, the temperature range of rapid temperature change test chambers is -55 to+80 ℃, while others are -40 to+80 ℃. There are two methods for determining the temperature change rate of the rapid temperature change test chamber: one is the average temperature rise and fall rate throughout the entire process, and the other is the linear temperature rise and fall rate (actually the average speed every 5 minutes). The average speed throughout the entire process refers to the ratio of the difference between the highest and lowest temperatures within the temperature range of the test chamber to the time. At present, the technical parameters of temperature change rate provided by various environmental testing equipment manufacturers abroad refer to the average rate throughout the entire process. The linear temperature rise and fall rate refers to the guaranteed temperature change rate within any 5-minute time period. In fact, for the rapid temperature change test chamber, the most difficult and critical stage to ensure the linear temperature rise and fall speed is the cooling rate that the test chamber can achieve during the last 5 minutes of the cooling period. From a certain perspective, the linear heating and cooling speed (average speed every 5 minutes) is more scientific. Therefore, it is best for the experimental equipment to have two parameters: the average temperature rise and fall speed throughout the entire process and the linear temperature rise and fall speed (average speed every 5 minutes). Generally speaking, the linear heating and cooling speed (average speed every 5 minutes) is half of the average heating and cooling speed throughout the entire process. 6. Wind speed According to relevant standards, the wind speed inside the temperature and humidity chamber during environmental testing should be less than 1.7m/s. For the test itself, the lower the wind speed, the better. If the wind speed is too high, it will accelerate the heat exchange between the surface of the test piece and the airflow inside the chamber, which is not conducive to the authenticity of the test. But in order to ensure uniformity within the testing chamber, it is necessary to have circulating air inside the testing chamber. However, for rapid temperature change test chambers and comprehensive environmental test chambers with multiple factors such as temperature, humidity, and vibration, in order to pursue the rate of temperature change, it is necessary to accelerate the flow velocity of the circulating airflow inside the chamber, usually at a speed of 2-3m/s. Therefore, the wind speed limit varies for different usage purposes. 7. Temperature fluctuation Temperature fluctuation is a relatively easy parameter to implement, and most test chambers produced by environmental testing equipment manufacturers can actually control temperature fluctuations within a range of ± 0.3 ℃. 8. Uniformity of temperature field In order to simulate the actual environmental conditions that products experience in nature more accurately, it is necessary to ensure that the surrounding area of the tested product is under the same temperature environment conditions during environmental testing. Therefore, it is necessary to limit the temperature gradient and temperature fluctuation inside the test chamber. In the General Principles of Environmental Test Methods for Military Equipment (GJB150.1-86) of the National Military Standard, it is clearly stipulated that "the temperature of the measurement system near the test sample should be within ± 2 ℃ of the test temperature, and its temperature should not exceed 1 ℃/m or the total maximum value should be 2.2 ℃ (when the test sample is not working). 9. Precision control of humidity The humidity measurement in the environmental testing chamber mostly adopts the dry wet bulb method. The manufacturing standard GB10586 for environmental testing equipment requires that the relative humidity deviation should be within ± 23% RH. To meet the requirements of humidity control accuracy, the temperature control accuracy of the humidity test chamber is relatively high, and the temperature fluctuation is generally less than ± 0.2 ℃. Otherwise, it will be difficult to meet the requirements for humidity control accuracy. 10. Cooling method selection If the test chamber is equipped with a refrigeration system, the refrigeration system needs to be cooled. There are two forms of test chambers: air-cooled and water-cooled.   Forced air cooling     Water-cooling Working conditions The equipment is easy to install, only need to power on. The ambient temperature should be lower than 28℃. If the ambient temperature is higher than 28℃, it has a certain impact on the refrigeration effect (preferably with air conditioning), the circulating cooling water system should be configured. Heat exchange effect  Poor (relative to the water-cooling mode)  Stable, good  Noise Large (relative to the water-cooling mode)    Less
    Read More
  • Maintenance of refrigeration compressor for constant temperature and humidity test chamber, cold and hot shock test chamber Maintenance of refrigeration compressor for constant temperature and humidity test chamber, cold and hot shock test chamber
    Oct 14, 2024
    Maintenance of refrigeration compressor for constant temperature and humidity test chamber, cold and hot shock test chamber Article summary: For environmental monitoring equipment, the only way to maintain long-term and stable use is to pay attention to maintenance in all aspects. Here, we will introduce the maintenance of the compressor, which is an important component of the constant temperature and humidity test chamber and the cold and hot shock test chamber Detailed content: Maintenance plan for refrigeration compressor: As the core component of the refrigeration system in the constant temperature and humidity test chamber, the maintenance of the compressor is essential. Guangdong Hongzhan Technology Co., Ltd. introduces the daily maintenance steps and precautions for the compressor in the constant temperature and humidity test chamber and the cold and hot shock test chamber 1、 Carefully check the sound of the cylinders and moving parts at all levels to determine if their working condition is normal. If any abnormal sound is found, immediately stop the machine for inspection; 2、 Pay attention to whether the indicated values of pressure gauges at all levels, pressure gauges on gas storage tanks and coolers, and lubricating oil pressure gauges are within the specified range; 3、 Check if the temperature and flow rate of the cooling water are normal; 4、 Check the supply of lubricating oil and the lubrication system of the moving mechanism (some compressors are equipped with organic glass baffles on the side of the cross head guide rail of the machine body), You can directly see the movement of the crosshead and the supply of lubricating oil; The cylinder and packing can be inspected for oil discharge using a one-way valve, which can check if the oil injector is inserted into the cylinder Oil injection situation; 5、 Observe whether the oil level in the body oil tank and the lubricating oil in the oil injector are below the scale line. If they are low, they should be refilled in a timely manner (if using a dipstick, stop and check); 6、 Check the temperature of the intake and exhaust valve covers at the cross guide rail of the crankcase with your hand to see if it is normal; 7、 Pay attention to the temperature rise of the motor, bearing temperature, and whether the readings on the voltmeter and ammeter are normal. The current should not exceed the rated current of the motor. If it exceeds the rated current, the cause should be identified or the machine should be stopped for inspection; 8、 Regularly check whether there are any debris or conductive objects inside the motor, whether the coil is damaged, and whether there is friction between the stator and rotor, otherwise the motor will burn out after starting; 9、 If it is a water-cooled compressor and water cannot be immediately supplied after the water is cut off, it is necessary to avoid cylinder cracking due to uneven heating and cooling. After parking in winter, the cooling water should be drained to prevent freezing and cracking of the cylinder and other parts; 10、 Check whether the compressor vibrates and whether the foundation screws are loose or detached; 11、 Check whether the pressure regulator or load regulator, safety valve, etc. are sensitive; 12、 Pay attention to the hygiene of the compressor, its associated equipment, and the environment; 13、 Gas storage tanks, coolers, and oil-water separators should regularly release oil and water; 14、 The lubricating machine used should be filtered by sedimentation. Differentiate the use of compressor oil between winter and summer
    Read More
  • EC-85EXT, Superior constant temperature bath (800L) Specifications EC-85EXT, Superior constant temperature bath (800L) Specifications
    Nov 13, 2014
    EC-85EXT, Superior constant temperature bath (800L)Specifications Project Type Series EXT Function Temperature occurs in a way Dry ball method Temperature range -70 ~ + 150 ℃ Temperature variation amplitude Below the + 100℃ ± 0.3 ℃ Above the + 101℃ ± 0.5 ℃ Temperature distribution Below the + 100℃ ± 0. 7 ℃ Above the + 101℃ ± 1.0 ℃ The temperature drops the time +125 ~-55 ℃ Within 36 minutes (10℃ / minute mean temperature change) Temperature rise time -55 ~+125 ℃ Within 36 minutes (10℃ / minute mean temperature change) The internal volume of the uterus was tested 800L Test room inch method (width, depth and height) 1000mm × 800mm × 1000mm Product inch method (width, depth and height) 1470mm × 2240mm × 2000mm Make the material  External outfit  Test room control panel machine room Cold interductile steel plate is dark gray Inside Stainless steel plate (SUS304,2B polished) Broken heat material Test room Hard synthetic resin glass wool door Hard synthetic resin foam cotton, glass cotton Project Type Series EXT Cooling dehumidifying device  Cooling-down method Mechanical section shrinkage and freezing mode and binary freezing mode  Cooling medium Single segment side R404A Binary high temperature / low temperature side R404A / R23 Cooling and dehumidifier Multi-channel mixed heat sink type The condenser (water-cooled) Calorifier Form Nickel-chromium heat-resistant alloy heater Blower Form Stir fan  Controllerl The temperature is set -72.0 ~ +152.0 ℃ Time setting Fanny 0 ~ 999 Time 59 minutes (programmed type)0 ~ 20000 Time 59 minutes (value) Set decomposition energy   Temperature  0.1 ℃ ,Time of 1 minute Indicate accuracy Temperature ± 0.8℃ (typ.), time ± 100 PPM Vacation type Value or program Stage number 20-stage / 1 program The number of procedures The maximum number of incoming force (RAM) programs is 32 programs The maximum number of internal ROM programs is 13 programs Round-trip number Maximum of 98 or unlimited Number of round-trip repeats Maximum 3 time Displace the end Pt 100Ω ( at 0 ℃ ),grade ( JIS C 1604-1997 ) Control action When splitting the PID action Endovirus function Early delivery function, standby function, setting value maintenance function, power outage protection function, Power action selection function, maintenance function, transportation round-trip function, Time delivery function, time signal output function, overrising and overcooling prevention function, Abnormal representation function, external alarm output function, setting paradigm representation function, Transport type selection function, the calculation time represents the function, the slot lamp lamp function Project Type Series EXT Control panel Equipment machine LCD operating panel (type contact panel), Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal, Time signal output terminal, power cord connector  Protective device Refrigerating cycle Overload protection device, high blocking device Calorifier Temperature over-rise protection device, temperature fuse Blower Overload protection device Control panel Leakage breaker for power supply, fuse (heater,), Fuse (for operating loop), temperature rise protection device (for testing), Temperature rise overcooling prevention device (test material, in microcomputer) Pay belongs to the product Test material shed shed by * 8 Stainless steel shed (2), shed receiving (4) Fuse Operating loop Protection Fuses (2) Operating specification (1 )  Else Bolus (Cable hole: 1) Equipment products Adventitia Hard borosilicate glass270mm× 190mm 1   Cable hole Bore size 50mm 1   The trough inside the lamp AC100V 15W White hot ball 1   Wheel   6   Horizontal adjustment   6   Electrovirus characteristics Source  * 5.1 AC Three-phase  380V  50Hz Maximum load current 60 A Capacity of the leakage breaker for the power supply 80A Sensory current  30mA Power distribution thickness 60mm2 Rubber insulation hose Coarseness of grounding wire 14mm2  Cooling water * 5.3 Water yield 5000 L /h (When the cooling water inlet temperature is 32℃) Water pressure 0.1 ~ 0.5MPa Side pipe diameter of the device PT1 1/4  Tubing  drain-pipe   * 5.4 PT1/2 Product weight 700kg  
    Read More
  • EC-85MHPM-W, High load corresponding constant temperature and humidity tank (800L) EC-85MHPM-W, High load corresponding constant temperature and humidity tank (800L)
    Nov 13, 2014
    EC-85MHPM-W, High load corresponding constant temperature and humidity tank (800L) Project Type Series MHPM-W Function Temperature and humidity mode The way of wet ball Temperature range -40 ~ + 100 ℃ Humidity range 20 ~ 98%RH (According to the anaphase 3 items) Changes in temperature and humidity ± 0.3 ℃ / ±2.5%RH Temperature and humidity distribution ± 0.5 ℃ / ±5.0%RH The temperature drops the time +20 ~ -40 ℃ Within 75 minutes Temperature rise time -40 ~ + 100 ℃ Within 50 minutes The internal volume of the uterus was tested 800 L Test room inch method (width, depth and height) 1000mm × 800mm × 1000mm Product inch method (width, depth and height) 1400mm × 1190mm × 1795mm Make the material External outfit Test room control panel machine room Cold steel plate, cold steel plate beige (Color table 2.5Y8 / 2) Inside Stainless steel plate (SUS304,2B polished) Broken heat material Test room Hard synthetic resin ― Door Hard synthetic resin foam cotton, glass cotton Project Type Series MHPM-W Cooling removal, wet device  Cooling-down method Mechanical section shrinkage mode  Cooling medium R404A The oneself can shrink machine Output (number of staff) 1.5kW (1) Cooling and dehumidifier Multi-channel mixed heat sink type The condenser Multi-channel mixed radiator sink (air cooling) Calorifier Form Nickel-chromium heat-resistant alloy heater Volume 3.5kW   Humidifier Form Steam generation Volume 1.8Kw×2 Blower Form Multi-channel mixed radiator sink (air cooling) Motor capacity 40W Feed water unit The water supply cylinder Water supply method Water quality Pure water * Automatic water supply ("Please refer to the automatic water supply.") Volume  Gravity type Moisturizing disk  Gravity type Controller Temperature setting range -42.0 ~ + 102.0 ℃ Humidity setting range 0 ~ 98%RH (Dry Bulb Temperature 10 ~ 85 ℃ ) Time set range 0 ~ 999Time of 59 min(Program setting type) 0 ~ 20000 Time of 59 min(The value type) Set decomposition energy Temperature 0.1℃, humidity 1% RH for 1 min Indicate accuracy Temperature ± 0.8℃ (tp.), humidity ± 1% RH (tp.), time ± 100 PPM Vacation type Value or program Stage number 20-stage / 1 program The number of procedures The maximum number of incoming force (RAM) programs is 32 programs The maximum number of internal ROM programs is 13 programs Round-trip number  98 times maximum or unlimited Number of round-trip repeats Maximum 3 heavy Displace the end Pt 100Ω ( at 0 ℃ ), grade B( JIS C 1604-1997 ) Control action When splitting the PID action Internal function Early delivery function, standby function, setting value maintenance function, power outage protection function, Power action selection function, maintenance function, transportation round-trip function, Time delivery function, time signal output function, overrising and overcooling prevention function, Abnormal representation function, external alarm output function, setting paradigm representation function, Transport type selection function, the calculation time represents the function, the slot lamp lamp function Project Type Series MHPM-W Control panel Equipment machine LCD operating panel (type contact panel), Represents lamp (power, transport, abnormal), test power supply terminal, external alarm terminal, Time signal output terminal, power cord connector  Protective device Refrigerating cycle Overload protection device, high blocking device Calorifier Temperature over-rise protection device, temperature fuse Humidifier Air burning prevention device, humidifying disc water level regulator Blower Overload protection device Control panel Leakage breaker for power supply, fuse (for heater, humidifier), Fuse (for operating loop), temperature rise protection device (for testing), Temperature rise overcooling prevention device (test material, in microcomputer) Offproducts (sets) House receiver (4), house board (2), wet ball wick (15), operation manual (1) Equipment products Adventitia Hard borosilicate glass  800mm× 800mm 2 Cable hole Bore size 50mm 1 The trough inside the lamp AC100V 15W White hot ball 2 Wheel   4 Horizontal adjustment   4 Electrovirus characteristics Source  AC   three-phase 380V 50Hz Maximum load current 25A Capacity of the leakage breaker for the power supply 50A Sensory current 30mA Power distribution thickness 14mm2 Rubber insulation hose Coarseness of grounding wire 5.5mm2 Tubing Drain-pipe PT1/2 Product quality 550kg
    Read More
  • ER-135MHP-W, Step in constant temperature and humidity bath, constant temperature and humidity chamber style book (water-cooled) ER-135MHP-W, Step in constant temperature and humidity bath, constant temperature and humidity chamber style book (water-cooled)
    Nov 10, 2014
    ER-135MHP-W, Step in constant temperature and humidity bath, constant temperature and humidity chamber style book (water-cooled) Project Type Form ER - 135MHP-W   Nature ability *1 *2 Temperature and humidity mode The way of wet ball Temperature and humidity model * 3 -40~+ 80 ℃ /10~95%RH Temperature and humidity electrode amplitude ± 0.3 ℃ / ± 2.5%RH Temperature and humidity distribution ± 0.75 ℃ / ± 5.0%RH The temperature drops the time +20~ -40 ℃ Within 170 minutes Temperature rise time +20~+ 80 ℃ Within 50 minutes Noise value *6 Room outside 65dB Indoor side 79B Statutory freezing capacity ( 50 / 60Hz ) 2.57/2.81 Product shape-inch method (width, depth and height) Please refer to the configuration diagram       Warehouse board laboratory Bed area 13 . 0 m2   The shape of inch method *7 Wide 3600mm Deep 3600mm High 2325mm Have oneself Intensification method *7 Wide 3450mm Deep 3450mm High 2100mm External materials Ivory painted steel plate Inside the material Stainless steel plate (SUS304) Broken heat material Hard synthetic resin The weight resistance of the board 5.9 k N/ m2 {600 kgf / m2 } Door (width and height) 830 mm × 1800 mm Single piece open Measurement (width and height) 190 mm × 320 mm The door Indoor light capacity (quantity) 60W( 2 individual ) Cable hole φ 50 1 (500mm from bed surface height, 1 side) Controller model and shape method Wide, deep and high EU - 6 5MH*2 unit 1200 mm × 710 mm × 1940 mm   External outfit  Indoor stainless steel sheet and plate (SUS304)  Outdoor L coating color on steel plate; ivory color: color table mark 2.5Y8/2 ) Operate Steel plate coating color; ivory color: color table mark2.5Y8/2   The oneself can shrink machine Cooling medium R404A Output of freezer 1 (staff) 2.2kW(1) × 2 Cooling and dehumidifier Multi-channel mixed heat sink type Condenser (staff number) Transverse fin type (2) 2 Calorifier Form Nickel-chromium heat-resistant alloy heater Volume 5 . 7kW(3 φ 200V) × 2 Humidifier Form SUS 316L Make a surface steam type humidifier Volume 5 . 4kW(3 φ 200V) × 2 Air supply device Form (quantity) Single air supply multi-blade air fan (1) Survival outside the machine 0 Pa{ 0 mmAq} Blowing rate 25 (m3 / min ) Back to the uterus 1100( r / min ) Motor form VTFO-K Motor output (pole number) 0.75kW(6) × 2 Moisturizing water supply installation  The water supply cylinder Water quality * 8 Electrical conductivity is below 10 μ S / cm Volume 4L × 2 Feedwater way Moisturizing disk  gravity type Wet ball Wick disc Electromagnetic valve type Controller The temperature is set -42.0 ~+ 82.0 ℃ Humidity is set 0~98% RH ( Dry Bulb Temperature 10 ℃ ~ 80 ℃ ) Time setting Fanny 0 ~ 999 Time 59 points (programmed type)0 ~ 20000 Time 59 points (value) Set decomposition energy Temperature 0.1 ℃ , humidity 1% RH , time 1 minute Indicate accuracy  Temperature  ± 0.8 ℃ ( typ. ), humidity ± 1% RH ( typ. ),  time ± 100 PPM Vacation type Value or program Stage number 20 stage / 1 form The number of procedures Number of incoming force (RAM) programs, maximum 32 programs / internal ROM programs, maximum 13 programs Round-trip number Maximum of 98 or unlimited Number of round-trip repeats Maximum 3 time Displace the end JP t 100Ω (a t 0 ℃ ), grade B(JIS C1604-1997 ) control action When splitting the PID action Endovirus function Early delivery function, standby function, setting value maintenance function, power outage protection function, Power action selection function, maintenance function, transportation round-trip function, Time delivery function, time signal output function, overrising and overcooling prevention function, Abnormal representation function, external alarm output function, setting paradigm representation function, Transport type selection function, the calculation time represents the function, the slot lamp lamp function Project Type Form ER - 55MHP-W Temperature control machine Control panel Equipment machine Color LCD operating panel (force contact mode), Represents the lamp (power supply, transport, abnormal), test material power supply system royal terminal, External alarm terminal, time signal output terminal, power cord connector,  Protective device Warehouse plate test uterine cavity Test material with Temperature rise overcooling prevention function (microcomputer: automatic setting) Temperature rise ter (set: Set above 60℃) Refrigerating cycle Overload protection device, high blocking device Calorifier Temperature over-rise protection device, temperature fuse   Humidifier Air burning prevention device (2 weight), wet plate water level regulator Blower Overload protection device Control panel Leakage breaker for power supply, fuse (for heater, humidifier), Fuse (for operating loop), fan reverse prevention relay In addition to control Frost system Defrosting method * 5 Stop cycling defrosting (freezer transport stop), Heat up defrosting Cold control But the system water   Concondensation control mode (quantity)   Water-making valve for cooling water (2) 2 Repeat measure Therstat (operated) 300kg × 2 Electricity Endyceps sinensis particular nature Source Communication  three phases 380V 50Hz Maximum load current 45A × 2 The power supply uses a leakage breaker 60A ( Sensory current 30mA) × 2 Power distribution Rubber insulation hose 22m m2 × 2 Coarseness of grounding wire 5.5m m2 × 2 cold step back water *9  Water yield;the yield of water 40L / min (  inlet water temperature  32 ℃ ) × 2 Water pressure 0.1~0.5M Pa{ 1.0~5.0 kgf / cm2 }  water temperature 18~ 32 ℃ Pipe-inch method Inlet / exit PT 1 parent seat / PT 1 parent seat Humidifying water supply pipe distribution inch method * 8 PT 1 / 2 parent seat arrange water Pipe-inch method drain-pipe PT 1 / 2 parent seat Its   Product products (staff)  reduce pressure(1) * 8 Wet ball wick (15) Take instructions (1)
    Read More
  • Road LED Text Reliability Test Road LED Text Reliability Test
    Oct 09, 2024
    Road LED Text Reliability Test Environmental resistance test: Vibration test, transportation package drop test, temperature cycle test, temperature and humidity test, impact test, waterproof test Durability test: High and low temperature preservation test, continuous switch operation test, continuous action test LED display reliability test conditions finishing: Vibration test: three-axis (XYZ) vibration, 10 minutes each, 10 ~ 35 ~ 10Hz sine wave, 300 ~ 1200 times/min, 3 minutes per cycle, vibration Fu 2mm Vibration tightening test: vibration + temperature (-10 ~ 60℃)+ voltage + load Drop test for transport packaging: Drop material slurry (at least 12mm thick), height depends on the purpose of use Temperature cycle: a. No boot test: 60℃/6 hours ← Rising and cooling for 30 minutes →-10℃/6 hours, 2cycle b. Boot test: 60℃/4 hours ← Rising and cooling 30 minutes →0℃/6 hours, 2cycle, power supply without packaging and load Temperature and humidity test: No power test: 60℃/95%R.H./48 hours Boot test: 60℃/95%R.H./24 hours/no packaging power supply load Impact test: impact distance 3m, slope 15 degrees, six sides Waterproof test: height 30 cm, 10 liters /min spray Angle 60 degrees, spraying position: front and back up, spraying range 1 square meter, spraying time 1 minute Humidity test: 40℃/90%R.H./8 hours ←→25℃/65%R.H./16 hours, 10cycle) High and low temperature preservation test: 60℃/95%R.H./72 hours →10℃/72 hours Continuous switch action test: Complete the switch within one second, shut down for at least three seconds, 2000 times, 45℃/80%R.H. Continuous action test: 40℃/85%R.H./72 hours/power on        
    Read More
  • Ac Solar Modules & Microinverters 1 Ac Solar Modules & Microinverters 1
    Oct 09, 2024
    Ac Solar Modules & Microinverters 1 The overall output power of the solar cell panel is greatly reduced, mainly because of some module damage (hail, wind pressure, wind vibration, snow pressure, lightning strike), local shadows, dirt, tilt Angle, orientation, different degrees of aging, small cracks... These problems will cause system configuration misalignment, resulting in reduced output efficiency defects, which are difficult to overcome traditional centralized inverters. Solar power generation cost ratio: module (40 ~ 50%), construction (20 ~ 30%), inverter (<10%), from the point of view of the cost proportion, the construction cost is as high as 1/3, if the inverter is directly installed on the module in production, the overall power generation cost can be greatly reduced. In order to overcome such problems, in 2008 developed a microinverter (microinverter) applied to the solar module, that is, each DC solar module is equipped with a direct conversion of direct current (DC) to AC (AC) small inverter, it can be directly installed behind the module or fixed frame, Through the micro inverter tracking, each module can operate at more than 95% of the highest power point (system more than 99.5% of the time is normal operation), such an advantage is for each module to optimize the output power, so that the entire solar power system output power to obtain the highest, for the design architecture, Even if some modules are covered by shadows, heat, dust... In addition, its power transmission value is connected to AC power supply, do not need complex and professional series and parallel, direct parallel output, can also reduce the attenuation between power transmission, recent research shows that the module assembly micro-inverter can increase the energy collection by 20%, a single module provides standard AC frequency power supply, Each module has arc protection, which can reduce the probability of arc occurrence. It can be seen that the failure rate of the centralized inverter is high, it must be replaced often, and its life is only about half of the module, if we use the micro inverter its output power is lower, it can improve the service life of the inverter. Since each module is behind the small inverter, the module does not need to configure another communication wire, can directly through the output wire of the AC Power supply, direct network communication, only need to install a power line network Bridge (Power line Ethernet Bridge) on the socket, do not need to set up another communication line, Users can directly access the web, iPhone, blackberry, tablet... Etc., watch the operation status of each module (power output, module temperature, fault message, module identification code), if there is an anomaly, it can be repaired or replaced immediately, so that the entire solar power system can operate smoothly. Output terminal of AC module: AC output, DC output, Control Interface Ac solar module English name: AC solar PV module ac pv module AC photovoltaic module AC Module PV systems composed of AC modules AC module-composed  PVAC Module Proprietary abbreviation: CVCF: constant voltage, constant frequency EIA(Energy Information Administration) The United States Energy Information Administration EMC: includes EMI(Electromagnetic interference) and EMS(electromagnetic tolerance) two parts EMI(Electromagnetic interference) : The electromagnetic noise generated by the machine itself in the process of performing the intended function is not conducive to other systems ETL: Electronic Testing Laboratory MFGR: Manufacturer HALT: Highly Accelerated Life Test. Halt: highly accelerated life test HAST(Highly Accelerated Stress Test) : Accelerated stress test HFRE: high frequency rectifier HFTR: high frequency transformer MEOST[Multiple Environment Over Stress Tests] : MEOST[multiple environment over stress tests] MIC(microinverter) : A microinverter Micro-inverters: indicates micro-inverters MPPT[Maximum Power Point Tracking] : indicates maximum power point tracking MTBF: mean time between failures NEC: National Electrical Code PVAC Module: AC solar module VVVF: Change voltage, change frequency            
    Read More
  • Ac Solar Modules & Microinverters 2 Ac Solar Modules & Microinverters 2
    Oct 08, 2024
    Ac Solar Modules & Microinverters 2 Ac module test specification: ETL Certification: UL 1741, CSA Standard 22.2, CSA Standard 22.2 No. 107.1-1, IEEE 1547, IEEE 929 PV Module: UL1703 Newsletter: 47CFR, Part 15, Class B Voltage Surge rating: IEEE 62.41 Class B National Electrical Code: NEC 1999-2008 Arc protection devices: IEEE 1547 Electromagnetic waves: BS EN 55022, FCC Class B per CISPR 22B, EMC 89/336/EEG, EN 50081-1, EN 61000-3-2, EN 50082-2, EN 60950 Micro-Inverter (Micro-inverter) : UL1741-calss A Typical component failure rate: MIL HB-217F Other specifications: IEC 503, IEC 62380 IEEE1547, IEEE929, IEEE-P929, IEEE SCC21, ANSI/NFPA-70 NEC690.2, NEC690.5, NEC690.6, NEC690.10, NEC690.11, NEC690.14, NEC690.17, NEC690.18, NEC690.64 Main specifications of AC solar module: Operating temperature: -20℃ ~ 46℃, -40℃ ~ 60℃, -40℃ ~ 65℃, -40℃ ~ 85℃, -20 ~ 90℃ Output voltage: 120/240V, 117V, 120/208V Output power frequency: 60Hz Advantages of AC modules: 1. Try to increase the power generation of each inverter power module and track the maximum power, because the maximum power point of a single component is tracked, the power generation of the photovoltaic system can be greatly improved, which can be increased by 25%. 2. By adjusting the voltage and current of each row of solar panels until all are balanced, so as to avoid system mismatch. 3. Each module has monitoring function to reduce the maintenance cost of the system and make the operation more stable and reliable. 4. The configuration is flexible, and the solar cell size can be installed in the household market according to the user's financial resources. 5. No high voltage, safer to use, easy to install, faster, low maintenance and installation cost, reduce the dependence on installation service providers, so that the solar power system can be installed by users themselves. 6. The cost is similar or even lower than that of centralized inverters. 7. Easy installation (installation time reduced by half). 8. Reduce procurement and installation costs. 9. Reduce the overall cost of solar power generation. 10. No special wiring and installation program. 11. The failure of a single AC module does not affect other modules or systems. 12. If the module is abnormal, the power switch can be automatically cut off. 13. Only a simple interrupt procedure is required for maintenance. 14. Can be installed in any direction and will not affect other modules in the system. 15. It can fill the entire setting space, as long as it is placed under it. 16. Reduce the bridge between DC line and cable. 17. Reduce DC connectors (DC connectors). 18. Reduce DC ground fault detection and set protection devices. 19. Reduce DC junction boxes. 20. Reduce the bypass diode of the solar module. 21. There is no need to purchase, install and maintain large inverters. 22. No need to buy batteries. 23. Each module is installed with anti-arc device, which meets the requirements of UL1741 specification. 24. The module communicates directly through the AC power output wire without setting up another communication line. 25. 40% less components.
    Read More
1 2
A total of2pages

leave a message

leave a message
If you are interested in our products and want to know more details,please leave a message here,we will reply you as soon as we can.
submit

home

products

WhatsApp

contact us